Note!
If you have selected Routine dialog and the dialog System
settings/Dialog options is disabled for routine dialog then you can
change back to expert dialog by entering Metrohm as the user in the
main dialog. If you are using login then a user who works with expert
dialog must log in.
The selection of method commands (list of commands) and the fixed
keys ([Print] and [Manual] only) which can be used can only be
opened when the system is being operated in expert dialog.
Open the Dialog options/Command list dialog.
Disable the checkboxes for the command groups which are not to
be used. These will then be shown in gray in the method editor for
the selection of the commands. These settings apply to both expert
dialog and routine dialog.
Open the Dialog options/Fixed keys dialog. Only the [Print]
and [Manual] (manual operating of the titration system) fixed keys
can be disabled.
Disable the checkboxes for the fixed keys which are not to be used.
These settings apply to both expert dialog and routine dialog.
3.7.4 Routine
dialog:
disabling functions
A standard configuration which is suitable for routine dialog is stored:
methods can be loaded but not altered or newly created. Determination
data (results) can be viewed and saved, but not modified. You can
adapt this standard configuration as follows:
Open Dialog options/Routine dialog in order to deactivate
those buttons which are to be disabled in the main dialog, the live
display and in manual control. With PC Control you can also disable
individual menus. This dialog can only be opened when the system
is being used in expert dialog.
Deactivate the checkboxes for the functions which are to be
disabled in routine dialog.
You can disable individual functions in the corresponding dialogs with