5.3 Determinations

726 Titroprocessor, Instructions for use

90

90

Statistics and sample series
In certain cases it may be a good idea to set the statistics limit to a lower

value than the total number of samples. If double or triple determinations

are to be carried out on the same sample within a sample series and

statistical data are to be calculated for each of them then the statistics limit

should be set to 2 or 3 respectively.

The statistical calculations are carried out for each sample but with these

settings the statistics are reset after each 2 or 3 determinations and the

working memory cleared. In this way statistical calculations are obtained for

double or triple determinations.
Warning! Statistical results are only calculated according to formulas which

have been defined as a 'Statistics formula' (formula type 'stat.') in the
formula editor. The statistics should be initialized before the start of a

sample series.