12.1 Error messages

726 Titroprocessor, Instructions for use

250

250

RECOGN 4:

enter value for EP-window "X" from...

RECOGN 5:

enter value for EP-window "X" ...to

RECOGN 6:

equal limits for EP-window "X"

Different values must be entered for the limits in EP window XX

RECOGN 7:

EP-window "X" and EP-window "Y" overlapping

Different EP windows must not overlap. Correct the values for the limits.

RECOGN 8:

EP-window "X" out of general limits

Correct the limits for EP window X, so that the EP window lies completely

within the general area or correct the general limits accordingly.

REPORTS

REPORTS 1:

no printer connected

Either no external printer has been connected or the internal printer has nor

been enabled. Check the configuration of the interface, see page 43ff.

REPORTS 2:

interface configured as COMM. and printer

An RS232 interface has been defined as the printer interface and also as the

communication interface. This can cause problems during data transfer.

Connect one of the instruments to the second RS232 interface.

REPORTS 3:

no COMM interface configured

For the required action the RS232 interface which is to be used for data

transfer must be defined as the communication interface. See page 43ff.

REPORTS 4:

too many variables. Output aborted

There are too many printer data to be printed out. The required reports can

be produced individually via the key.

RS232

RS232 1:

interface X possibly defect

Problems have been found with RS232 interface X which indicate a hard-

ware fault. Please contact the service department of your local Metrohm

agency.

RS232 2 ... 13: interface X: ...

The Titroprocessor cannot establish a stable, serial connection via RS232

interface X. Check whether the transfer parameters of the Titroprocessor

and those of the connected instrument are identical. Initialize the interface

again.

Is an instrument actually connected? Is the correct cable being used?