beautypg.com

Metrohm 726 Titroprocessor User Manual

Page 118

background image

7.6 KFT

Karl Fischer titrations

726 Titroprocessor, Instructions for use

112

112

What to do if ...

Observation

Possible causes and remedies

Final addition lasts too

long and

increments are too

small.

"Will never end!"

Increase 'Min.increment".

Alter switch-off criterion, e.g. try increasing the stop drift or use a shorter

switch-off time as the stop criterion.

Change the solvent for problematic samples, e.g. for ketones or

aldehydes the use of 2-methoxyethanol or a mixture of methanol/glacial

acetic acid for amines can be recommended, see technical publications.

The increments at the

end of the titration

are too large.

"Overshoots".

Reduce 'Max. Rate'.

An indication of the optimal Max. Rate is provided by the following

experiment: observe the drift during conditioning and add the sample

substance without starting the titration.

Select a value below the highest drift as the 'Max. Rate'.

Optimize the arrangement of the electrode and buret tip and improve the

stirring.

At the end of the

titration the

solution is too

brown.

The methanol content in the solvent solution is too low. Change the

solvent.

The electrode may be coated; wipe off with acetone.

Solution becomes

darker after each

titration.

Renew solvent.

The electrode may be coated; wipe off with acetone.

The drift increases

after every titration.

Does the sample only release water slowly? Use the KF oven.

Are acids esterified in the sample? Change the solution more frequently.

Increase the buffering capacity of the solvent.

Does the sample contain ketones or aldehydes? Use special reagents

which are suitable for ketones and aldehydes.

The endpoint is

reached "too quickly".

Reduce Max.rate.

The titration times

become longer and

longer.

With 2-component reagents the buffering capacity of the solvent may be

exhausted. Change the solution.

If the drift increases at the same time please see above.