8. Command reference

8. Command reference

726 Titroprocessor, Instructions for use

169

169

fer to the numerical measured value alterations. The control

direction can also be determined automatically. The control

direction is then obtained from the initial measured value

and the preset target endpoint.

End-point

Target endpoint

for KFC_IPOL*

-2000…250…2000 mV

for KFC_UPOL*

-200 …25…200 µA

The target endpoint defines the measured value which the

titration control is aiming for. The standard settings (250 mV

or 25 µA) are ideal in most cases and should only be al-

tered in exceptional cases. See also page 110ff.

Adjustment

Predefined parameter sets

slow | medium | fast

A selection can be made from three different sets of pa-

rameters for the simplified setting of the titration parame-

ters. Each of these sets includes proven controller settings

which are suitable for most requirements.

slow

for low water content or samples which only

release their moisture slowly

medium standard parameters for unproblematic

titrations, optimized for the commonest

applications

fast

for high water content of non-critical

samples

If the titration parameters have to be adapted for specific ti-

tration problems the different control parameters can be

altered individually.

The [Custom adjust] softkey opens a further dialog
window in which user-defined settings can be carried out

(see page 164).

Fill. rate

Buret filling rate

0.01 mL/min…150 mL/min | max.

At the end of a titration the buret is automatically filled. The

maximum value for the filling rate depends on the cylinder

size of the buret used.

Buret cylinder max. speed

1 mL

3 mL/min

2 mL

6 mL/min

5 mL

15 mL/min

10 mL

30 mL/min

20 mL

60 mL/min

50 mL

150 mL/min