8. Command reference

8. Command reference

726 Titroprocessor, Instructions for use

187

187

Special beaker

Name of special beaker

8 characters

The special beaker can be specified by its name. The en-

try field contains a selection list in which first the rack and

then the required special beaker can be selected.

to tower

Rack rotation target

1 | 2

The tower in front of which the special beaker is to be po-

sitioned can be freely selected.

LIFT

Positioning lift

The lift with the titration head can be positioned to the nearest millimeter. It

is recommended that predefined lift positions are used and fixed in the con-

figuration for the specific rack.

Command Parameters Parameters

ДДДДДДДДДДДДД ДДДДДДДДДДДДДДДДДДДДДДДД ДДДДДДДДДДДДДДДДДДДДДДДД

LIFT

to tower 1 to position 0 mm

on tower

Address of the lift

1 | 2

In a 2-tower changer the lifts can be selected at will. The

lift address always refer to the active changer.

to position

Target position of the lift

work | shift. | rinse. | rest | spec | 0…325 mm

The user-defined lift positions (work position, shift posi-

tion, rinsing position, special position) can be defined at

will for each rack, depending on the beaker sizes used.

work position

lift position in which e.g. a titration can

be carried out

shift position

lift position in which the rack can be

freely rotated without damaging the

sensors or other items.

rinsing position

lift position for rinsing the titration

head

rest position

start position at the upper end of

the tower (0 mm)

special position freely selectable lift position