Metrohm 726 Titroprocessor User Manual
Page 193
8. Command reference
8. Command reference
726 Titroprocessor, Instructions for use
187
187
Special beaker
Name of special beaker
8 characters
The special beaker can be specified by its name. The en-
try field contains a selection list in which first the rack and
then the required special beaker can be selected.
to tower
Rack rotation target
1 | 2
The tower in front of which the special beaker is to be po-
sitioned can be freely selected.
LIFT
Positioning lift
The lift with the titration head can be positioned to the nearest millimeter. It
is recommended that predefined lift positions are used and fixed in the con-
figuration for the specific rack.
Command Parameters Parameters
ДДДДДДДДДДДДД ДДДДДДДДДДДДДДДДДДДДДДДД ДДДДДДДДДДДДДДДДДДДДДДДД
LIFT
to tower 1 to position 0 mm
on tower
Address of the lift
1 | 2
In a 2-tower changer the lifts can be selected at will. The
lift address always refer to the active changer.
to position
Target position of the lift
work | shift. | rinse. | rest | spec | 0…325 mm
The user-defined lift positions (work position, shift posi-
tion, rinsing position, special position) can be defined at
will for each rack, depending on the beaker sizes used.
work position
lift position in which e.g. a titration can
be carried out
shift position
lift position in which the rack can be
freely rotated without damaging the
sensors or other items.
rinsing position
lift position for rinsing the titration
head
rest position
start position at the upper end of
the tower (0 mm)
special position freely selectable lift position