4. Manual operation
Manual operation
726 Titroprocessor, Instructions for use
67
67
Softkeys
[Select
determ.]
opens a selection list of the determinations in the working
memory. A different determinationfor printing out reports
can be selected.
[Select
report]
if a titration or measurement within a determination has
been selected then this softkey opens the selection list
containing possible report modes , see below.
[Global
Reports]
opens the dialog window for selection of the global
reports, see below.
ЪДДД
Print Determ.report
ДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДД¿
³ ЪДДД
Print Global reports
ДДДДДДДДД¿ ³
³ ³ ³ ³
³ ³ Configuration ³ ³
³ ³ Data card ³ ³
³ ³ Internal Storage ³ ³
³ ³ Commands ³ ³
³ ³ Method ³ ³
³ ³ Sample Silo ³ ³
³ ³ Devices & Manual Control ³ ³
³ ³ Common Variables ³ ³
³ ³ Full statistics ³ ³
³ ³ Short statistics ³ ³
³ ³ GLP ³ ³
³ АДДДДДДДДДДДДДДДДДДДДДДДДДДД QUIT ДЩ ³
АДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДД QUIT ÄÙ
[Send]
starts the output of the selected report to the
communication interface. The report is transferred to the
instrument (e.g. a personal computer) which is connected
to the defined RS232 interface.
[Print]
starts the output of the selected report at the configured
printer.
4.7.2 General reports
General reports are not dependent on a determination. They present the
momentary settings effective on the instrument.
Configuration
Report of the settings in the configuration window and its
sub-windows
Data card
List of stored files on the data card
Internal
storage
List of stored files in the internal memory
Commands
List of command sequences of the current method,
without parameters
Method
Comprehensive listing of the current method , including
application remarks, command sequences and all their
parameters.
Sample Silo
List of all sample data in the sample silo