8. Command reference

8. Command reference

726 Titroprocessor, Instructions for use

151

151

Derivation

Recording the 1st derivative of the titration curve

on | off

If the derivative of the titration curve is to be shown then

derivation recording must be switched on.

U(pol)

Polarization voltage (only with SET_UPOL*)

-1270…400…1270 mV

The polarization voltage is the voltage which is applied to a

polarized electrode during amperometric titration. The re-

sultant measuring quantity is µA.

I(pol)

Polarization current (only with SET_IPOL*)

-127…1…127 µA

The polarization current is the current which is applied to a

polarized electrode during voltametric titration. The resul-

tant measuring quantity is mV.

Stop conditions

The stop conditions correspond to those of the KFT mode, see page 163f.

8.4.2 Custom Adjustments

Custom adjustments allow the control parameters of the titration to be finely

tuned. Starting with a set of parameters (fine, medium, coarse) after press-

ing the softkey [User defin.] under 'Adjustments' on the parameter
page the individual parameters can be altered in the dialog window 'Custom

adjustments'.

ЪДДД

Custom adjustments

ДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДД¿

³ ³

³ Dynamics pH 2.00 ³

³ Min. rate 25.00 æL/min Stop crit. drift ³

³ Max. rate 10.00 mL/min Stop drift 20 æL/min ³

³ ³

АДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДД QUIT ÄÙ

Dynamics

Measurement range for controlled titration

for SET_PH:

0.1…220 | off

for SET_U, SET_IPOL:

1…100.0…2000 mV | off

for SET_UPOL:

0.1…10.0…200.0 µA | off

The control range defines the measurement range in

which controlled titration is carried out before the given

endpoint. Outside this range addition is made at the

maximum dosing rate. For steep curves a sufficiently

large control range should be selected; for flatter curves

this can be smaller. If the control range is switched off