Metrohm 726 Titroprocessor User Manual
Page 259
12
12. . Troubleshooting
Troubleshooting
726 Titroprocessor, Instructions for use
253
253
STORAGE 10:
terminate "KeyCTRL mode" first
For the required action the key simulation mode 'keyCTRL' of the RS232 in-
terface must be switched off. This can be carried out under [Config.]
[Interfaces].
STORAGE 11:
"RS 232 Interface" still active
There is still a connection to an external instrument via the RS232 interface.
This data communication should first be terminated.
STORAGE 12:
invalid file name
The file name is not in accordance with the conventions for the naming of
files, see page 73.
STORAGE 13:
RS232 interface, check settings
The interface parameters do not meet the requirements of the selected ac-
tion. Correct the settings under [Config.] [Interfaces].
STORAGE 14:
copy process aborted
A copying process could not be completed owing to an operator action. The
selected file(s) have not been copied.
STORAGE 15:
unexpected file block size
STORAGE 16:
memory block size violation
Faults have occurred during file transfer. Check the interface parameters of
the PC software.
STORAGE 17:
undefined error
A fault has occurred which cannot be specified more closely. Please con-
tact the service department of your local Metrohm agency.
STORAGE 18:
invalid type of file
The selected file cannot be stored in the given memory area as it does not
meet the specified type.
STORAGE 19:
invalid file
The selected file is damaged or of an unsuitable type.
STORAGE 20:
remove balance from interface X
No balance can be active on the same interface during file transfer. Switch it
off or connect it to a different interface.