8.10 Sample changer commands

726 Titroprocessor, Instructions for use

184

184

when processing the CHANGER command the sample
changer will be initialized again, the rack code of the

sample rack used will be read and compared with the

configuration data (and thus the rack name) in the rack

tables.

Lift rate 1 / 2

Lift speed at tower 1 and 2

3…25 mm/s

If a lift head is heavily loaded with titration accessories,

several sensors, etc. then the lift speed should be re-

duced. It is possible to set different speeds for the two

lift towers.

Shift rate

Rotational speed of the turntable

3…20 °/s

The rotational speed of the sample changer turntable

can be varied according to the load.

Beaker test mode

Test mode of the beaker sensor

single |both

After each MOVE command of the sample changer the

beaker sensor of the changer tower checks whether a

beaker is present in the rack position which has been

moved to.

For racks whose arrangement of beaker positions does

not correspond with the setting angle of both the

changer towers (e.g. 14-place, 16-place racks) the test

mode 'single' should be switched on. In this way only

the beaker sensor of the approached tower is activated.

If parallel processing is carried out at two towers then,

provided a suitable rack is used, the test mode 'both'

can be selected.

Shift direction

Rotational direction of the sample rack

auto | + | –

There are different settings for the rotational direction of

the sample rack:

auto

the changer always selects the

shortest rotational path

+

counterclockwise direction, towards

higher rack positions

clockwise, towards lower rack

positions

Take care that samples which have not been processed

are not contaminated by dripping sensors or similar.