12.1 Error messages

726 Titroprocessor, Instructions for use

254

254

SYSTEM

SYSTEM 1:

Execution not possible while measuring

The selected function cannot be carried out while a method is running. Wait

until the method has ended.

SYSTEM 2:

Measuring start inhibited at this state

It is not possible to start a method while alterations are being carried out in

the method editor at the same time. Close the method editor.

TITRATION

TITRATION 1:

dosing drive not responding

The connection to the selected dosing drive has been interrupted during the

titration. Stop the determination, switch off the Titroprocessor and reconnect

the dosing drive. Switch the Titroprocessor on again.

TITRATION 2:

A/D converter error

The analog/digital converter of the sensor interface is faulty. Please contact

the service department of your local Metrohm agency.

TITRATION 3:

measuring point list overflow

More than 1000 measuring points have been recorded. An evaluation of the

further measuring points is not possible. Correct the control parameters of

the titration, see page 136 (DET), 144 (MET), 151 (SET), 158 (KFT), or 175

(MEAS).

TITRATION 4:

end-point list overflow

More than 9 equivalence points have been found. Further equivalence

points cannot be evaluated. Optimize the recognition parameters and, if

necessary, the control parameters of the titration, see page 138.

TITRATION 5:

max. number of measuring points reached

More than 1000 measuring points have been recorded. An evaluation of the

further measuring points is not possible. Correct the control parameters of

the titration, see page 136 (DET), 144 (MET), 151 (SET), 158 (KFT), or 175

(MEAS).

TITRATION 6:

end-point already reached

The set target end-point has already been reached at the start of the titra-

tion. An attempt may have been made to retitrate a sample which has al-

ready been titrated.

TITRATION 7:

no end-point set

No set target end-point has been entered for an end-point titration (SET,

KFT, SEC, KFC). Add the set target end-point to the method.