8. Command reference

8. Command reference

726 Titroprocessor, Instructions for use

189

189

Stirring rate

Stirrer speed

1…15 | off

The stirring speed of each individual stirrer can be varied

in 15 steps.

RESET_C

Initializing changer

After sample racks have been changed it is necessary to initialize the sam-

ple changer. This is done by rotating the sample turntable to the basic posi-

tion and reading off the magnetic code for identification of the rack. This is

normally carried out manually, but can also be programmed in a method

(preferably in the start sequence, see OMOVE below).

Command Parameters Parameters

ДДДДДДДДДДДДД ДДДДДДДДДДДДДДДДДДДДДДДД ДДДДДДДДДДДДДДДДДДДДДДДД

RESET_C

Selection 1

Selection

Changer address

1 | 2

Changer 1 or changer 2 are defined by the setting of the

EBus address selector disk on the sample changer, see

page 24.

(OMOVE

Start sequence for a series of samples

)OMOVE

When a series of samples is being processed it is often necessary to carry

out certain working steps a single time before the sample series itself. For

example, these could be rinsing procedures or conditioning steps for sen-

sors.

The definition of the necessary command sequence is made in an OMOVE

sequence (OMOVE = Opening Moves), which is enclosed by the starting

marker (OMOVE and the end marker )OMOVE.

Example: Rinsing a sensor:

t/s Command Parameters Parameters

ДДДДД ДДДДДДДДДДДДД ДДДДДДДДДДДДДДДДДДДДДДДД ДДДДДДДДДДДДДДДДДДДДДДДД

(OMOVE

CHANGER Selection 1 Rack name

Lift rate 1 25 Lift rate 2 25

Shift rate 20 Beaker test mode single

Shift direction auto On beaker error MOVE_S

MOVE_B Special beaker waste to tower 1

LIFT on tower 1 to position rinse mm

PUMP Address 1.rinse State on

10 PUMP Address 1.aspir State on

PUMP Address 1.rinse State off

PUMP Address 1.aspir State off

LIFT on tower 1 to position shift mm

)OMOVE

...