beautypg.com

Metrohm 726 Titroprocessor User Manual

Page 264

background image

12.2 Remedying storage errors

726 Titroprocessor, Instructions for use

258

258

Warning:

Only switch off the 726 Titroprocessor in the ready condition (* ready *),

never during a running determination or saving process.

When the instrument is switched on various error messages can appear.

The following message appears if data from one determination only are af-

fected:

WARN. DETSTORE: Determination x.x corrupt. Clear working memory ?

In this case the error can be remedied in the following way:

WARN. DETSTORE:

Determination 1.1 corrupt.

Clear working memory ?

Working memory

is cleared.

Corrupted Determi-
Bestimmung

nation is cleared.

Y

N

Main
page

Power

on

Determination

defect

If the query is answered with then only the data in the defect determi-

nation will be deleted; this is normally sufficient. If the Titroprocessor should

then behave irregularly, delete the working memory by initializing the stati-

stics function. Enter

Statistics 'rst'

on the main page.

If a serious memory error occurs the following error message appears:

WARN. DETSTORE: Corrupt memory detected. continue anyway ?

This query can be answered with , but the Titroprocessor may then be-

have irregularly. You have the possibility of saving your data.

If the query is answered with , then proceed as follows: