8.10 Sample changer commands

726 Titroprocessor, Instructions for use

186

186

Value

Numerical value of alteration or allocation

0…1…999

The numerical value of how much the SAMPLE variable

has been altered or what value it should be allocated

can be freely selected (whole numbers).

MOVE_S

Position current sample beaker

The command MOVE_S can be used to position the current sample beaker

in front of tower 1 or 2 of the sample changer.

The rack is rotated so that the beaker position of the rack, which corre-

sponds to the contents of the SAMPLE variable (see above) comes to rest

in front of the tower given in the command. Rotation of the rack is only pos-

sible when the lift (or both lifts) are located on or above its rotation position

(see LIFT command below).

After the racks has been rotated the corresponding beaker sensor checks

whether a beaker is present in this position. If this is not the case then the

procedure defined in the beaker test mode (see page 184) will be carried

out.

Command Parameters Parameters

ДДДДДДДДДДДДД ДДДДДДДДДДДДДДДДДДДДДДДД ДДДДДДДДДДДДДДДДДДДДДДДД

MOVE_S

Sample to Tower 1

Sample to tower

Rack rotation target

1 | 2

With a 2-tower changer the target of the rack rotation

and thus the positioning of the current sample beaker

can be selected at will. In this way it is possible to sub-

ject the same sample to different processing steps at

two work stations during a method run. For example,

these could be two different titrations or more compli-

cated sample preparation steps.

MOVE_B

Positioning special beaker

Up to eight special beakers (rinsing, conditioning beakers, etc.) with perma-

nently assigned rack positions can be defined for a single sample rack. The

command MOVE_B can be used in a method run to position a special

beaker in front of either of the changer towers.

Rotation of the rack is only possible when the lift (or both lifts) are located on

or above its rotation position (see LIFT command below).

Command Parameters Parameters

ДДДДДДДДДДДДД ДДДДДДДДДДДДДДДДДДДДДДДД ДДДДДДДДДДДДДДДДДДДДДДДД

MOVE_B

Special beaker to tower 1