beautypg.com

Metrohm 726 Titroprocessor User Manual

Page 254

background image

12.1 Error messages

726 Titroprocessor, Instructions for use

248

248

MONITORING 4:

auto save aborted; XXX bytes needed

There is too little space available in the target memory. Delete files which are

no longer required.

MONITORING 7:

new method while statistics ON

A new method has been loaded or the existing method has been altered

while statistics is switched on. During the statistical calculation this could

lead to results with the same name which have been produced in different

ways being calculated with each other.

MONITORING 8:

working memory full; delete determinations

There is not enough working memory space available for the next determi-

nation. Delete determinations from the working memory (in the file man-

ager). Also reduce the set value for the statistics counter, if this is neces-

sary. This value lays down how many determinations are to be held in the

working memory before the working memory is cleared (during the next de-

termination).

MONITORING 10: method autoload disabled: quit method editor

While alterations are being made in the method editor it is not possible to

load a method automatically, as can be defined for a sample series in the

sample silo. Close the method window with in order to allow the

sample series to proceed.

MONITORING 11: GLP data not saved, irregular termination

A GLP validation has not been concluded properly.

MONITORING 12: check GLP data

The validation interval for a validation method has expired. Carry out a new

validation.

MONITORING 13: Do you want to reset all statistics data?

Query before resetting the statistics. The working memory will be cleared.

MONITORING 14: method line XX : no condition!

The method contains a CASE or EXIT command without any condition.

Check the method line 'xx'.

MONITORING 15: Impossible to execute "SHOW" while editing

Because of a SHOW command in the current method a particular dialog

page is to be shown; this is not possible while an entry field is being edited.

Close the editing mode. The required page will be shown after the next

SHOW command.