12. . Troubleshooting
Troubleshooting
726 Titroprocessor, Instructions for use
231
231
ARITHMETIC
ARITHMETIC 1:
unexpected arithmetic error, code: XXXX
An arithmetical error has occurred in an internal calculation while a method
is being carried out. Check the current method.
BALANCE
BALANCE 1:
invalid sample size
An invalid weight has been received. Check the balance settings.
BALANCE 6:
data for current sample are not admitted
Invalid sample data have been received. Check the balance settings.
CAL DATA
CAL DATA 1:
calibration interval expired, sensor X
The sensor mentioned should be subjected to a new sensor test. See page
275.
CAL_PARAM
CAL_PARAM 1:
buffer list full
No further buffer can be defined. Delete unnecessary buffers from the list.
CHANGER
CHANGER 1:
no sample changer detected
No sample changer has been connected.
CHANGER 2:
changer 'X' max. powerlimit reached
The sample changer is overloaded or blocked. Check the sample changer.
CHANGER 3:
no rack data found on changer 'X'
The magnet code of the rack placed on the sample changer cannot be as-
signed to any rack table. Define the corresponding rack data with the ac-
companying rack code under [Config.] [Sample changer].