12. . Troubleshooting

Troubleshooting

726 Titroprocessor, Instructions for use

231

231

ARITHMETIC

ARITHMETIC 1:

unexpected arithmetic error, code: XXXX

An arithmetical error has occurred in an internal calculation while a method

is being carried out. Check the current method.

BALANCE

BALANCE 1:

invalid sample size

An invalid weight has been received. Check the balance settings.

BALANCE 6:

data for current sample are not admitted

Invalid sample data have been received. Check the balance settings.

CAL DATA

CAL DATA 1:

calibration interval expired, sensor X

The sensor mentioned should be subjected to a new sensor test. See page

275.

CAL_PARAM

CAL_PARAM 1:

buffer list full

No further buffer can be defined. Delete unnecessary buffers from the list.

CHANGER

CHANGER 1:

no sample changer detected

No sample changer has been connected.

CHANGER 2:

changer 'X' max. powerlimit reached

The sample changer is overloaded or blocked. Check the sample changer.

CHANGER 3:

no rack data found on changer 'X'

The magnet code of the rack placed on the sample changer cannot be as-

signed to any rack table. Define the corresponding rack data with the ac-

companying rack code under [Config.] [Sample changer].