12. . Troubleshooting
Troubleshooting
726 Titroprocessor, Instructions for use
237
237
DETSTORE
DETSTORE 1:
Corrupt memory detected. Continue anyway ?
The last determination was not properly concluded, probably because the
instrument was switched off during a storage process. In order to clean up
the resulting memory conflict the error message must be answered with
other memory areas. If the error message is answered with
Titroprocessor will start without taking any further measures. However, it
must be expected that the instrument may behave in an irregular manner
because of the memory conflict mentioned above.
DETSTORE 2:
Determination X.Y corrupt. Clear working memory?
The last determination was not properly concluded, probably because the
instrument was switched off. In order to clean up the resulting memory con-
flict the error message must be answered with
memory of all the files it contains. If
determination will be deleted.
DETSTORE 3:
Do you want to clear working memory ?
Because of memory conflicts (see previous error messages) it is recom-
mended that the working memory is cleared. This will be carried out if the
error message is answered with
lost. If
dertaken.
DETSTORE 4:
Storage initializing recommended. Initialize partially?
In order to remedy memory conflicts (see previous error messages) the
whole operator memory area (current methods, working memory, internal
memory) can be cleared by pressing
fected. Answering with
memory can be carried out (see next error message).
DETSTORE 5:
Total initialisation: restart and press
In order to remedy existing memory conflicts which could influence the
functioning of the Titroprocessor it may be necessary to carry out a total ini-
tialization of the instrument. After pressing the
stopped. Total initialization of the Titroprocessor is triggered by switching it
off and then on again with pressed-down space bar: this deletes all data
stored in the instrument (determinations, methods, GLP data, configuration
settings, etc.) and returns it to the standard settings.
Switching on without pressing down the space bar produces a normal start-
up; no data are deleted or altered.
DETSTORE 6:
lock impossible, quit dialog page
Running or reloading a method cannot be implemented as long as the cur-
rently opened dialog page has not been closed.
Close it and then press the