Metrohm 726 Titroprocessor User Manual
Page 283
13. Appendix
726 Titroprocessor, Instructions for use
277
277
The following limits can be defined:
Slope
Electrode slope
Zero point (mV)
Asymmetry potential at pH 7 (U
as
)
Streaming potential
Potential difference between stirred – unstirred
buffer
Response time
Maximum time for measured value setting, up to
1 mV of effective measured value
Test interval
Interval until next sensor test
By pressing the [Cancel] softkey alterations in the limits can be canceled.
Carry out sensor test
Enter a name for the sensor(s) together with its/their manufacturing and or-
der number . Set the cursor to the line containing the sensor to be tested.
Press the [Sensor test] key.
ЪДДД GLP ДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДД¿
³ ³
³ Method Validations Date Interval Result ³
³ ДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДД ³
³ЪДДД
Sensors
ДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДД¿ ³
³³ ³ ³
³³ SensoÚÄÄÄ
Sensor test
ДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДД¿ity ³ ³
³³ ДДДДД³ ³ДДД ³ ³
³³ 1.comb.³ Sensor comb. pH II ³ ³ ³
³³ 2.comb.³ ДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДД ³ ³ ³
³³ 3. ³ Meas. input A1 Temperature 25.0 §C ³ ³ ³
³³ 4. ³ Buffer type Metrohm Stirrer Int. A ³ ³ ³
³³ 5. ³ ³ ³ ³
³³ 6. ³ ³ ³ ³
³³ 7. ³ Prepare buffer solution pH 7.00 and pH 4.00 ³ ³ ³
³³ 8. ³ ³ ³ ³
³³ 9. ³ Press softkey [Start test] to proceed ³ ³ ³
³³10. АДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДД QUIT ДЩ ³ ³
³³ ³ ³
³³ ³ ³
³АДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДД QUIT ДЩ ³
³ ³
³ ³
АДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДД QUIT ДЩ
ЪДДДДДДДВДДДДДДДВДДДДДДД¿ ЪДДДДДДДВДДДДДДДВДДДДДДД¿ ЪДДДДДДДВДДДДДДДВДДДДДДД¿
³ Start ³ ³ ³ ³ ³ ³ ³ ³ ³ ³ ³
³ test ³ ³ ³ ³ ³ ³ ³ ³ ³ ³ ³
АДДДДДДДБДДДДДДДБДДДДДДДЩ АДДДДДДДБДДДДДДДБДДДДДДДЩ АДДДДДДДБДДДДДДДБДДДДДДДЩ
The sensor must be connected to measuring input A1, A2, B1 or B2. A
magnetic or rod stirrer should be connected to stirrer output 1 (Sensors A)
or 2 (Sensors B) on the rear panel of the Titroprocessor. The stirrer is con-
trolled automatically. Two beakers with buffer solutions pH 7 and pH 4 must
be prepared. The buffer type can be entered in the dialog window ‘Sensor
test’
. The measuring temperature can be measured automatically with a
pH electrode with a built-in Pt100 or Pt1000 temperature sensor or with a
separate temperature sensor. The measuring temperature is taken into ac-
count in the sensor test.
After pressing the [Start test] softkey the test proceeds automatically.