13. Appendix

726 Titroprocessor, Instructions for use

277

277

The following limits can be defined:

Slope

Electrode slope

Zero point (mV)

Asymmetry potential at pH 7 (U

as

)

Streaming potential

Potential difference between stirred – unstirred

buffer

Response time

Maximum time for measured value setting, up to

1 mV of effective measured value

Test interval

Interval until next sensor test

By pressing the [Cancel] softkey alterations in the limits can be canceled.

Carry out sensor test

Enter a name for the sensor(s) together with its/their manufacturing and or-

der number . Set the cursor to the line containing the sensor to be tested.

Press the [Sensor test] key.

ЪДДД GLP ДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДД¿

³ ³

³ Method Validations Date Interval Result ³

³ ДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДД ³

³ЪДДД

Sensors

ДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДД¿ ³

³³ ³ ³

³³ SensoÚÄÄÄ

Sensor test

ДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДД¿ity ³ ³

³³ ДДДДД³ ³ДДД ³ ³

³³ 1.comb.³ Sensor comb. pH II ³ ³ ³

³³ 2.comb.³ ДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДД ³ ³ ³

³³ 3. ³ Meas. input A1 Temperature 25.0 §C ³ ³ ³

³³ 4. ³ Buffer type Metrohm Stirrer Int. A ³ ³ ³

³³ 5. ³ ³ ³ ³

³³ 6. ³ ³ ³ ³

³³ 7. ³ Prepare buffer solution pH 7.00 and pH 4.00 ³ ³ ³

³³ 8. ³ ³ ³ ³

³³ 9. ³ Press softkey [Start test] to proceed ³ ³ ³

³³10. АДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДД QUIT ДЩ ³ ³

³³ ³ ³

³³ ³ ³

³АДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДД QUIT ДЩ ³

³ ³

³ ³

АДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДД QUIT ДЩ

ЪДДДДДДДВДДДДДДДВДДДДДДД¿ ЪДДДДДДДВДДДДДДДВДДДДДДД¿ ЪДДДДДДДВДДДДДДДВДДДДДДД¿

³ Start ³ ³ ³ ³ ³ ³ ³ ³ ³ ³ ³

³ test ³ ³ ³ ³ ³ ³ ³ ³ ³ ³ ³

АДДДДДДДБДДДДДДДБДДДДДДДЩ АДДДДДДДБДДДДДДДБДДДДДДДЩ АДДДДДДДБДДДДДДДБДДДДДДДЩ

The sensor must be connected to measuring input A1, A2, B1 or B2. A

magnetic or rod stirrer should be connected to stirrer output 1 (Sensors A)

or 2 (Sensors B) on the rear panel of the Titroprocessor. The stirrer is con-

trolled automatically. Two beakers with buffer solutions pH 7 and pH 4 must

be prepared. The buffer type can be entered in the dialog window ‘Sensor
test’

. The measuring temperature can be measured automatically with a

pH electrode with a built-in Pt100 or Pt1000 temperature sensor or with a

separate temperature sensor. The measuring temperature is taken into ac-

count in the sensor test.
After pressing the [Start test] softkey the test proceeds automatically.