13.5 Validation / GLP

The Titroprocessor 726 supports the monitoring of maintenance intervals as

well as system validation and sensor tests.

13.5.1 Maintenance intervals

The maintenance of the Titroprocessor 726 should be carried out at least

once per year by a service technician from your local Metrohm agency.

The service technician updates the service data and enters a suitable

maintenance interval.

When this maintenance interval has elapsed the Titroprocessor 726 displays

a warning when it is switched on and recommends that the operator should

contact the Metrohm service department.

The service data can be seen on the GLP page. They are called up by

pressing the key on the keyboard.

ЪДДД GLP ДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДД¿

³ ³

³ Method Validations Date Interval Result ³

³ ДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДД ³

³ 1. ³

³ 2. ³

³ 3. ³

³ 4. ³

³ 5. ³

³ ³

³ ³

³ ³

³ ³

³ Instrument ident. : 01106 ³

³ Last servicing : 1997-11-25 ³

³ done by : B. Miller ³

³ Program version : 5.726.0012 ³

³ ³

³ ³

³ ³

³ ³

³ ³

³ ³

АДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДД QUIT ДЩ

ЪДДДДДДДВДДДДДДДВДДДДДДД¿ ЪДДДДДДДВДДДДДДДВДДДДДДД¿ ЪДДДДДДДВДДДДДДДВДДДДДДД¿

³ ³ ³Delete ³ ³Define ³Extend.³Start ³ ³ ³Sample ³Sensors³

³ ³ ³method ³ ³method ³info ³valid. ³ ³ ³silo ³ ³

АДДДДДДДБДДДДДДДБДДДДДДДЩ АДДДДДДДБДДДДДДДБДДДДДДДЩ АДДДДДДДБДДДДДДДБДДДДДДДЩ

The service data are updated by the service technician.

13.5.2 Sensor test

The Titroprocessor 726 offers a sensor test procedure which checks

whether the pH electrode is still functioning properly. In order to guarantee

the suitability of the sensor for a specific (or general) application, separate

limits can be defined for each sensor; these must not be exceeded in the

sensor test.

The sensor test consists of an extended calibration with automatic control of

the stirrer 'Int.A' or 'Int.B'. The test can be carried out with any buffer set

containing the buffers pH 4 and pH 7. We recommend Metrohm buffers, but

other buffer sets can also be used.