8.8 MEAS

726 Titroprocessor, Instructions for use

176

176

ЪДДД

MEAS_PH* Parameters

ДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДД¿

³ ³

³ Data acquisition ³

³ ³

³ Meas. procedure multiple ³

³ Measuring interval off s Start val. pH off ³

³ Meas. cycles 1 Stop val. pH off ³

³ ³

³ ³

АДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДД QUIT ÄÙ

Data acquisition

Depending on the selected measuring procedure, data acquisition may

have different settings.

Meas. procedure

Single or multiple measurement

standard | multiple

Switches between standard and multiple measurement.

Meas. interval

Time interval between single measurements

0.1…9999 s | off

The measuring interval is the time pattern at which a

single measurement is triggered within the series. If a

drift-controlled measurement is recorded then this is not

accepted in the measuring point list immediately, but

only at the time at which the measure acceptance con-

ditions (drift) have been fulfilled. This can lead to an ir-

regular time pattern in the measuring point list.

If 'off

off

' is selected the single measurements will be made

at the shortest possible time interval (<50 ms).

Meas. cycles

Number of single measurements in a series

1…1000 | inf. | max

The maximum number of single measurements is 1000.

With 'inf.' the measuring series will not end, but after
1000 measuring points have been recorded no further

values will be entered in the measuring point list. With

'max' the series is stopped after 1000 measuring points.

Start val.

Measured value at the start of the series

for MEAS_PH*

pH -20…20 | off

for MEAS_U*, MEAS_IPOL*

-2000…2000 mV | off

for MEAS_UPOL*

-200 …200 µA | off

for MEAS_T*

-170…500°C | off

If a start value is set this means that the measuring se-

ries will only be started when the defined value is ex-

ceeded in the sample solution. The direction in which

the measurements alter is irrelevant.