Metrohm 726 Titroprocessor User Manual
Page 240
12.1 Error messages
726 Titroprocessor, Instructions for use
234
234
CHANGER 29:
changer 'X' error; Key[Reset]-->MAN CONTROL
There is a sample changer fault. It is necessary to reset the sample
changer. Press the
changer] and [Reset].
CHANGER 42:
missing type def. for rack no. "X"
A rack has been defined, but the type of rack has not been entered. Enter
the correct rack type (e.g. M12-0 for a 12-position Metrohm rack) for the
corresponding sample rack.
CHANGER 43:
same code for rack no. "X" and no. "Y"
Rack codes must be unambiguous. Alter the code of one of the racks men-
tioned. Note that the code must always coincide with the physical arrange-
ment of the magnet code of the racks.
CONFIG
CONFIG 1:
"standard file name" too long
The standard file name entered is too long for the type of file name. Shorten
the text to 5 characters.
CONFIG 2:
invalid password
Access to the access control page is protected by a password. Enter the
correct password.
CURVE
CURVE 1:
too few measuring points
In order to produce a measuring curve or a calibration curve at least 4
measurements must be recorded.
CURVE 2:
curve display not possible
Data required for the representation of a curve are not present. Make sure
that the measurement time recording function has been switched on if the
curve is to be produced on the time axis: see page 134.
CURVE 3:
missing scale parameters for X-axis
CURVE 4:
missing scale parameters for Y-axis
All the scale parameters have not been entered.
CURVE 5:
incomplete parameters
Not all the parameters required for the production of a curve have been en-
tered.