MALLOC

MALLOC 1:

last powerdown left inconsist. status. Continue anyway?

Switching off the instrument has caused memory conflicts. It is recom-

mended that the error message is answered with and that a part ini-

tialization of the user memory area is carried out, see error message

DETSTORE 4, page 237. If the error message is answered with the Ti-

troprocessor starts up without taking any further measures. However, it must

be expected that the instrument may behave in an irregular manner be-

cause of the memory conflict mentioned above.

MALLOC 3:

out of user memory

The working memory and the internal memory cannot accept any more

data. Delete determinations or methods which are no longer required from

these memories.

MALLOC 4:

out of system memory

The internal system memory is overloaded by printing jobs or other internal

functions. Terminate any printing jobs with the key combination

. Avoid several large printing jobs in sequence.

MALLOC 5:

reallocation of user memory failed

A serious fault has occurred during internal memory operations. Switch the

instrument off and then on again. If this error message appears again then a

total initialization of the whole memory is necessary which will delete all data

in the Titroprocessor. In order to trigger this switch the instrument on with

the space bar pressed down.

MALLOC 6:

reallocation of system memory failed

A serious fault has occurred during internal memory operations. Switch the

instrument off and then on again. If this error message appears again then a

total initialization of the whole memory is necessary which will delete all data

in the Titroprocessor. In order to trigger this switch the instrument on with

the space bar pressed down.

MAN_CAL

MAN_CAL 1:

invalid measured value

The measured value is not valid. Check the sensor and the buffer solution.

MAN_CAL 2:

temperature value invalid

The value given by the temperature sensor is invalid. Check it.

MAN_CAL 3:

value not in buffer table

A buffer value has been found which does not match any buffer solution of

the given manufacturer or type. Check the buffer parameters and the buffer

solution.