7. Titration and measuring modes

726 Titroprocessor, Instructions for use

121

121

IM

Initial measured value (unit depends on mode)

SM

Measured value after the start phase of a titration or measure-

ment

CM

Measured value at end of a titration or measurement

SV

Added starting volume in mL

CV

Added volume at end of a titration in mL

C

Concentration of reagent (value without units)

HNV#

HNP volume in mL (half neutralization potential)

HNP#

HNP value (unit depends on mode)

AP

Asymmetry pH (pH

as

) of the calibrated sensor used

SL

Slope of the calibrated sensor used

DRC

Volume drift in conditioned status (in µL/min)

DOD

Dosing duration in s, length of the dosing increments added

together (for SET and KFT modes)

ITD

Duration between end of conditioning and start of the

following titration in s

$RN

Sample number, run number

$AS

Autostart counter

$CRF

Conditioning flag (conditioning OK, yes/no (1/0)

COV

Conditioning volume in mL

$BC

Buffer counter

$LV

Live volume in mL, added volume to live value display

$LM

Live value, actual value to live value display

$LD

Live measurement duration in s, actual duration to live value

display

7.10.2 Mathematical functions

+

Addition

Subtraction

*

Multiplication

/

Division

^

Exponentiation

( )

Expression in brackets

LOG( )

Decimal logarithm

EXP( )

Decimal exponent

LN( )

Natural logarithm

SQRT( )

Square root

TST( , )

Test function (tested variable, substitution term)

ABS()

Absolute value (unsigned)

All the above mathematical functions can be used in formulas. Expressions

in brackets can be combined at will. The recognized rules of priority for

mathematical functions apply. The formula editor has an automatic syntax

checking function.
If errors such as missing or invalid variable contents occur while the calcula-

tion formulas are being carried out (at the conclusion of the titration or

measuring mode) then a warning is given and the error message is entered

in the error list on the result page. The error list is included in the result re-

port.