12.1 Error messages

726 Titroprocessor, Instructions for use

232

232

CHANGER 4:

no rack code found for changer 'X'

No magnet code can be read in at the sample changer. Use a rack with a

valid rack code.

CHANGER 5:

changer 'X' not ready

The sample changer is busy. Wait until it is ready and try the command

again.

CHANGER 7:

changer 'X' not accessible

The selected sample changer does not reply. Check the sampler connec-

tion 'External Bus' and the cable.

CHANGER 8:

changer 'X' already locked

The selected sample changer is busy processing a series of samples. It

cannot be accessed by manual operation.

CHANGER 9:

changer 'x' MOVE disabled, raise lift

The MOVE command cannot be carried out because the/a lift is located

below its shifting position. Raise the lift concerned to a safe position.

CHANGER 10:

changer 'X', unable to move lift

The lift is blocked mechanically. Ensure that the lift can be moved and try

again.

CHANGER 11:

changer 'X' no MOVE executed

The preconditions for a MOVE_S command have not been fulfilled. Check

whether the position to be moved to has not been defined as special beaker

(press [Config.] [Sample changer] select rack [Special beakers]).

CHANGER 12:

changer 'X', drive overload

The sample changer turntable is blocked. Ensure that the rack can be

turned and try again.

CHANGER 13:

changer 'X' invalid drive

CHANGER 14:

changer 'X' common access failure YYYY

There are connection problems between the sample changer and Titroproc-

essor. Please contact the service department of your local Metrohm agency.

CHANGER 15:

changer 'X', EBus timeout

The connection between the Titroprocessor and sample changer has been

interrupted. Check the 'External-Bus' connection socket.