8. Command reference

8. Command reference

726 Titroprocessor, Instructions for use

175

175

Data acquisition

Depending on the selected measuring procedure, data acquisition may

have different settings.

Meas. procedure

Single or multiple measurement

standard | multiple

Switches between standard and multiple measurement.

Sampling time

Time interval for raw data acquisition

for 'Standard' measuring procedure

1…9999 s | off

The sampling time determines the time pattern for pre-

senting the raw data which are to be taken into account

when determining the drift value during a drift-controlled

single measurement. It has no influence on the effective

measurement. The raw data can be shown as a curve.

This can be useful in assessing the setting behavior of a

sensor.

U(pol)

Polarization voltage (for MEAS_UPOL*)

-1270…400…1270 mV

The polarization voltage is the voltage which is applied to a

polarized electrode during amperometric measurement.

The resultant measuring quantity is I

pol

/µA.

I(pol)

Polarization current (for MEAS_IPOL*)

-127…1…127 µA

The polarization current is the current which is applied to a

polarized electrode during voltametric measurement. The

resultant measuring quantity is U

pol

/mV.

8.8.2 Multiple measurement

A multiple measurement is a series of (drift-controlled) single measure-

ments, each of which is made in a particular time pattern according the set

measuring parameters and produces a single measured value. Multiple

measurements can be used to record alterations of certain measuring

quantities such as pH or temperature over a longer period of time and show

them as a curve. This ensures that each individual measurement, thanks to

the drift-controlled measurement acceptance, is always a reliable one. Up to

1000 measured values can be recorded in this way.

The measuring procedure (standard measurement / multiple measurement)

can be selected under the extended parameters (softkey [*Param]) to-
gether with further data recording parameters.