13.1

13.1 Diagnosis

Diagnosis

726 Titroprocessor, Instructions for use

260

260

13 Appendix

13.1 Diagnosis

13.1.1 General

The 726 Titroprocessor is a very precise and dependable instrument.

Thanks to its rugged construction, it is highly unlikely that external mechani-

cal or electrical influences will have any adverse effect on its functions.
Although a fault in the instrument can not be excluded with certainty, the

possibility is greater that malfunctions are caused by wrong operation or

handling, through improper connections and the operation with third-party

devices.
Whatever the case, it is always advisable to localise the fault with the diag-

nostic tests, which can be performed quickly and simply. The customer

need call Metrohm service only when the instrument really has a fault. Fur-

ther, he can use the results of the specific diagnostic function to provide the

service engineer with much more precise information.
In the case of inquiries, always quote the serial number on model plate (see

Figure.page 4), the program version (see configuration, page 33) and if ap-

plicable the error message.

Procedure

Procedure

The diagnostic menu listed in section 13.1.2

13.1.2 shows all components for

which detailed instructions (diagnostic steps) are available for checking the

functionality.
In the case of a possible malfunction, we advise you to perform either the

corresponding diagnostic step or all diagnostic steps as a routine check on

the instrument.
The reactions of the instruments to the instructions must be compared with

the descriptions in the diagnostic step. If the instruments do not show the

expected reaction ("No" case), the appropriate diagnostic step must be re-

peated to exclude operating errors. However, it is highly probable that re-

peated wrong reactions indicate a malfunction.

Equipment required:

Equipment required:

3.496.8510

test plug to check the Remote interface

3.496.8480

test plug to check the RS 232 interface

3.496.8570

test plug to check the Barcode interface

1.767.0010

simulator to check the analog inputs