8.7 KFC

726 Titroprocessor, Instructions for use

168

168

KFC modes produce the following system variables:

COV

conditioning volume; reagent consumption for conditioning

DRC

volume drift in conditioned state (µL/min)

$CRF

conditioning flag; correctly conditioned yes/no (1 or 0)

These system variables can be used to apply an automatic drift volume re-

duction:

Formula:

Drift red.

R1=DRC/60*(ITD+MD)/1000

mL

EP corr.

R2=EP1-R1

mL

Here it is assumed that owing to external influences a volume drift is effec-

tive during the sample titration and whose value corresponds to the volume

drift in a conditioned state (variable DRC). Its duration is equivalent to the

period between conditioning switch-off and sample titration switch-off

(ITD+MD * ).

* ITD= Intermediate titration duration, time between conditioning switch-

off and the start of the next titration in s.

MD= Mode duration, sample titration time in s.

Assignment parameters for KFC modes:

The assignment parameters (Comments, Reagent … etc.) for the KFC

modes are the same as those of the DET modes, see page 129f.

8.7.1 Specific parameters for KFC modes [ * Param ]

The specific parameters for the respective KFC mode can be found under

the softkey [* Param ].

ЪДДД

KFC_IPOL* Parameter

ДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДД¿

³ ³

³ Titration conditions ³

³ Direction neg. Temperature 25.0 øC ³
³ End-point at 250 mV I(pol) 50 µA ³

³ Adjustment medium ³

³ ³

³ Fill. rate max mL/Min ³

АДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДД QUIT ÄÙ

Titration conditions

These are the actual titration parameters which control the measurement re-

cording and the dosing behavior during the titration itself.

Direction

Control direction

pos. | neg. |auto

The direction of the measurement alterations is important

for titration control in the KFC mode. Normally the 'neg.'
setting should be used. The 'pos.' and 'neg.' settings re-