beautypg.com

Metrohm 726 Titroprocessor User Manual

Page 188

background image

8.9 CAL

726 Titroprocessor, Instructions for use

182

182

The simplest form of an automatic calibration loop could look like this:

Command Parameters Parameters

ДДДДДДДДДДДДД ДДДДДДДДДДДДДДДДДДДДДДДД ДДДДДДДДДДДДДДДДДДДДДДДД

(CAL

(CASE $BC = 1

MOVE_B Special beaker Buffer 7 to Tower 1

CASE $BC = 2

MOVE_B Special beaker Buffer 4 to Tower 1

)CASE

LIFT on Tower 1 to position work mm

STIR_C Address 1 Stirring rate 6

CAL* pH Sensor calibration

Sensor comb. glass Signal drift 2 mV/min

Meas. input A1 Equilibr.time auto s

Temperature 25.0 øC

STIR_C Address 1 Stirring rate off

LIFT on Tower 1 to position shift mm

...

)CAL

The precondition for this type of automatic calibration is that the special

beaker positions (with the corresponding names) of the sample rack used

have been correctly configured.

On the flash card supplied with the 726 Titroprocessor you will find a

method (SC_CAL.mth) in which the automatic calibration is completely

programmed as the start sequence for a series of titrations. You only need

to complete it with a titration mode and to enter the buffers to be used (in

CAL* mode). Do not forget to define the required special beakers for the

sample rack to be used.

Behavior if calibration problems occur

If during an automatic calibration a calibration error occurs and the run is

not interrupted the following automatic procedures are provided:

If a buffer cannot be properly measured then this buffer will not be taken

into account for the calibration. The calibration will nevertheless be con-

tinued. A corresponding error message will appear in the result report.

If at least two buffers can be properly measured then these values will be

used for the calibration. The method will be continued.

If only one buffer or no buffer at all can be properly measured the cali-

bration will be interrupted with an error message. You must then decide

whether the method should nevertheless be continued or terminated. In

the first case standard calibration data (pH

as

7.000, slope 1.000) will be

assumed.