2 dosing devices, 1 automatic reagent check, 2 dosing device functions – Metrohm 726 Titroprocessor User Manual

Page 59: 3 manual dosing

background image

4. Manual operation

Manual operation

726 Titroprocessor, Instructions for use

53

53

B1, B2

Measuring electrode of measuring group

‘Sensors B’

Bdiff

Differential amplification arrangement at

measuring group ‘Sensors B’

BT

Temperature sensor Pt100/1000 of

measuring group ‘Sensors B’

Bpol

Polarized electrode of measuring group

‘Sensors B’

In order to operate an instrument or a sensor move the field cursor to the

corresponding editing block and press the corresponding softkey. Press the

or key to change between dosing devices and sensors.

In order to address additional instruments, e.g. dosing devices B1 … C4 or

a sensor in measuring group B, press the key.

с

> or

can be used to switch between the screens of the various interfaces

or measuring groups.

4.2 Dosing devices

Twelve dosing devices can be connected simultaneously to the 726

Titroprocessor and comfortably operated by 'Manual control'. It is possible

to operate dosing devices manually while determinations are being carried

out. The dosing devices in use are blocked for manual operation by the

Titroprocessor; they are marked as 'locked'.

The name and concentration of a reagent can be entered for each dosing

device. The operator can use this to force the Titroprocessor to check the

reagent required for a method before it is used.

4.2.1 Automatic reagent check

In order to ensure that the dosing device settings in a method coincide with

the current settings of the manual control page (in which the actual

occupancy of the dosing devices should be entered) it is possible to force a

comparative check to be carried out.

If a reagent name and/or a concentration for a dosing device have been

entered in the method and in the 'Manual control page' then an automatic

reagent check is carried out. The two entries are compared with each other.

Variations in a reagent entry or its concentration produce an error message

at the start of the particular command or mode.

The automatic reagent check is suppressed if no entries have been made

for reagent and concentration in the method or if the corresponding entries

have been completely deleted in 'Manual control page'.

³ Dos. drive A1 700

ß Manual control page

³ Reagent AgNO3

³ Concentr. 0.01 mol/L

â Method

DET_PH* Determination of chloride

Reagent AgNO3 Sensor

Concentr. 0.01 mol/L Meas. input A1

Dos. drive A1

In titration modes the 'Reagent' field is a selection list in which the
reagent names and concentrations available on the manual control page

are present.