6. Sample data and sample silo
726 Titroprocessor, Instructions for use
97
97
Data output in the current sample data
The output pointer (<) marks the silo line which will be used for the next
determination in a sample series. The output marker is set in the same way
as the input marker (Softkey [<--Out pointer]).
In order to process a sample series the sample data must first be entered in
the sample silo. Place the output marker on the silo line with which the
sample series is to start.
The number of samples or silo lines to be processed can be set in the
status window (
'Auto start'
. Not only a number can be entered. The entry 'Silo' has
the effect that all silo lines after the output marker will be processed.
The method which is to be used for one or more particular samples can be
defined in the method input field of each silo line. The specified method is
loaded into the working memory before the affected line(s) are processed
and carried out with the sample data of the current silo line. If this possibility
is used then no alterations can be made to the particular methods loaded
while the sample series is being processed. An open method editor
prevents another method from being loaded.
If no method has been specified then the current method in the working
memory will be used. When a determination has been completed the
sample data remain in the sample silo but the output marker is moved
downwards by one position.