5.3 Determinations

726 Titroprocessor, Instructions for use

82

82

Destination

memory area for automatic storage of
determinations, see above.

Determ.name

the file name of the current determination appears
here.

Auto start

single determinations or automatic sample series;
set 'off' for single determinations.

Silo

on/off switch for sample silo; set 'off' for single
determinations.

Changer

address of sample changer used; no entry is
required for single determinations.

Now prepare the sample.

Enter the sample data. This can be done directly in the main page of the

Titroprocessor or by calling up the sample window with the

key.

ЪДДД

Current Sample

ДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДД¿

³ ³

³ Ident1 Ident2 Ident3 Size Unit ³

³ Sample ³

³ ³

³ Remark ³

³ ³

АДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДД QUIT ÄÙ

The sample data can be entered manually. It is also possible to accept

the sample weight and units directly from a balance .

Sample data can also be entered while a determination is being carried

out ; however, this must be done before the calculations have been

completed, i.e. before the titration or measurement has been completed

in which the sample weight is to be used for calculation.

For automatic data transfer from a balance it is not absolutely necessary

that the main page or the sample window is visible. The incoming

sample data are received by the RS232 interface, interpreted and filed

on the main page or – if switched on – in the sample silo. Even after the

determination has been carried out it is still possible to enter new sample

data for the recalculation of the results.

If sensors and dosing devices have been correctly connected (see

installation pages 22 and 25) the dosing devices can now be prepared for

the determination.

To profit from the full accuracy of your dosing devices, cylinders and

tubings must be rinsed before each use or at least once per day. This

eliminates gas bubbles that may form in the dosing cylinder if the Dosino is

not used for a longer time interval.

Use the PREP function regularly.

SAMPLE