8. Command reference

8. Command reference

726 Titroprocessor, Instructions for use

155

155

* ITD= Intermediate titration duration, time between conditioning switch-

off and the start of the next titration in s.

MD= Mode duration, sample titration time in s.

Assignment parameters for SEC modes:

The assignment parameters (Comments, Reagent … etc.) for the SEC

modes are the same as those of the DET modes, see page 129f.

8.5.1 Specific parameters for SEC modes [ * Param ]

The specific parameters for the respective SEC mode can be found under

the softkey [* Param ].

ЪДДД

SEC_PH* Parametere

ДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДД¿

³ ³

³ Titration conditions ³

³ Direction auto Temperature 25.0 øC ³

³ End-point pH ³

³ Adjustment medium ³

³ ³

³ Fill. rate max mL/min ³

АДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДД QUIT ÄÙ

Titration conditions

These are the actual titration parameters which control the measurement re-

cording and the dosing behavior during the titration itself.

Direction

Control direction

pos. | neg. | auto

The control direction must be entered for titration control in

the SEC mode. Normally the 'auto' setting should be used.
The control direction is then determined from the initial

measured value and preset target endpoint. The control

direction can also be fixed in advance. The possible set-

tings 'pos.' and 'neg.' refer to the numerical measured
value alterations.

End-point

Target endpoint

for SEC_PH*

pH -20…20

for SEC_U*, SEC_IPOL*

-2000…2000 mV

for SEC_UPOL*

-200 …200 µA

The target endpoint defines the measured value which the

conditioning control is aiming for. Correct calibration of the

sensor used is highly recommended.