8. Command reference
8. Command reference
726 Titroprocessor, Instructions for use
167
167
8.7 KFC
Conditioning for a Karl Fischer titration is implemented as an independent
mode. The KFC mode has its own set of parameters. The settings for con-
ditioning and the titration itself should, if possible, not differ too much from
each other.
A KFC mode does not produce a measuring point list and therefore no
curves can be shown. No facilities for calculations are provided.
Conditioning is normally carried out in the background, i.e. as soon as the
conditioning mode has been started the 726 Titroprocessor can carry out
other commands (e.g. CRTL_RS1, etc.). A conditioning process also re-
mains active even when a determination has been completed, provided that
it has not been explicitly terminated, see below.
A conditioning mode carried out in the background is controlled by the
WAIT_C command (see page208f). This checks whether the conditioned
has been achieved and, if necessary, switches the conditioning mode off.
Example:
KFC_U*
KFC_U* ДДДДДДДДДДДДД
ДДДДДДДДДДДДД¿
¿
CTRL_RS1 ... |
CTRL_RS1 ... |
... KFC_U*
... KFC_U*
... ...
... ...
WAIT_C ...
WAIT_C ... ДДДДДДДДД
ДДДДДДДДДЩ
Щ
ASK
ASK
KFT_U*
KFT_U*
The WAIT_C command synchro-
nizes the background conditioning
process with the method being car-
ried out in the foreground. The
WAIT_C command interrupts the
regular method run until the back-
ground process (conditioning) has
been completed.
Note! Every titration, measuring or calibration mode also terminates back-
ground conditioning. Achievement of the conditioned state is not taken into
account.
Two KFC modes are available for conditioning to a preset endpoint; these
differ in the measured quantity or measuring mode:
KFC_UPOL*
measured quantity µA (selectable pol.voltage in mV)
KFC_IPOL*
measured quantity mV (selectable pol.current in µA)
The following information applies to all KFC modes. Any variations are
noted.
KFC_IPOL*
conditioning
for Karl Fischer titration
Reagent Sensor
Concentr. Meas. input Apol
Dos. drive A1
applies to
⇒
KFC_UPOL*, KFC_IPOL*