8. Command reference

8. Command reference

726 Titroprocessor, Instructions for use

167

167

8.7 KFC

Conditioning for a Karl Fischer titration is implemented as an independent

mode. The KFC mode has its own set of parameters. The settings for con-

ditioning and the titration itself should, if possible, not differ too much from

each other.

A KFC mode does not produce a measuring point list and therefore no

curves can be shown. No facilities for calculations are provided.

Conditioning is normally carried out in the background, i.e. as soon as the

conditioning mode has been started the 726 Titroprocessor can carry out

other commands (e.g. CRTL_RS1, etc.). A conditioning process also re-

mains active even when a determination has been completed, provided that

it has not been explicitly terminated, see below.

A conditioning mode carried out in the background is controlled by the

WAIT_C command (see page208f). This checks whether the conditioned

has been achieved and, if necessary, switches the conditioning mode off.

Example:

KFC_U*

KFC_U* ДДДДДДДДДДДДД

ДДДДДДДДДДДДД¿

¿

CTRL_RS1 ... |

CTRL_RS1 ... |
... KFC_U*

... KFC_U*
... ...

... ...
WAIT_C ...

WAIT_C ... ДДДДДДДДД

ДДДДДДДДДЩ

Щ

ASK

ASK
KFT_U*

KFT_U*

The WAIT_C command synchro-

nizes the background conditioning

process with the method being car-

ried out in the foreground. The

WAIT_C command interrupts the

regular method run until the back-

ground process (conditioning) has

been completed.

Note! Every titration, measuring or calibration mode also terminates back-

ground conditioning. Achievement of the conditioned state is not taken into

account.

Two KFC modes are available for conditioning to a preset endpoint; these

differ in the measured quantity or measuring mode:

KFC_UPOL*

measured quantity µA (selectable pol.voltage in mV)

KFC_IPOL*

measured quantity mV (selectable pol.current in µA)

The following information applies to all KFC modes. Any variations are

noted.

KFC_IPOL*

conditioning

for Karl Fischer titration

Reagent Sensor

Concentr. Meas. input Apol
Dos. drive A1

applies to

KFC_UPOL*, KFC_IPOL*