8.13 Communication functions

726 Titroprocessor, Instructions for use

202

202

Pattern

Bit pattern for status of lines

8 characters (0 | 1 | *)

The 8 or 14 output lines of the remote socket can be set

at will. This is done by setting an 8 or 14-place bit pattern,

in which each bit is assigned to an output line.

Output 13 12 11 10 9 8 7 6 5 4 3 2 1 0

Bit 13 12 11 10 9 8 7 6 5 4 3 2 1 0

(Bits are always numbered from right to left)

Example:

CTRL_Rm Remote A Pattern ************1

sets the output line 0 to active (=set), which, for example,

if a Metrohm Ion meter is connected would have the ef-

fect of a start command.

It is recommended that input lines which are not relevant

should be masked with an asterisk (*) so that their status

is not altered.

A line set with a CTRL command (1) remains statically

active. Under some circumstances it may be necessary to

reset it at a certain time (0).

It is of course possible to set several output lines at the

same time with a more complex bit pattern (e.g.
*****1****0001

).

8.13.3 RS 232 interfaces

The RS 232 interfaces of the 726 Titroprocessor together with the corre-

sponding communication commands allow very flexible solutions for de-

manding automation applications to be programmed. For example, as well

as data communication with computers, Metrohm instruments which under-

stand the Metrohm remote language can be remotely controlled by the Ti-

troprocessor to their full extent; various functions of a modern analytical bal-

ance can also be used such as text display of the balance display, etc.

Just as with remote interfaces it is possible both to transmit data (character

sequences) and also to interpret incoming character sequences (scanning).

When the RS 232 interfaces are used it is an absolute necessity that the

transmission parameters of the instruments concerned are the same (see

page 44). When the data communication commands CTRL_RSx and

SCAN_RSx are used one of the communications interfaces should be se-

lected and the data communication mode set to 'LIMS'. The use of the
SCAN_RSx command requires that the data receive mode of the particular

interface is switched on. No other instruments such as printers, etc., should

0 = inactive (high)

1 = active (low)

* = no change