8. Command reference

8. Command reference

726 Titroprocessor, Instructions for use

195

195

Command Parameters Parameters

ДДДДДДДДДДДДД ДДДДДДДДДДДДДДДДДДДДДДДД ДДДДДДДДДДДДДДДДДДДДДДДД

(CASE Condition

... ...

CASE Condition

... ...

EXIT Condition

... ...

CASE Condition

... ...

)CASE

A CASE sequence is processed so that all CASE conditions are individu-

ally checked step by step in sequence. As soon as one of the conditions is

applicable all the sequence of commands following this CASE command

will be carried out up to the next CASE command. After this has been proc-

essed the CASE sequence will be exited without checking the other condi-

tions.

A conditional premature exit from the CASE sequence can be forced with

the EXIT command, for which an analogous condition may also be given.

The command process will be continued with the first command line after

the )CASE end marker.

If none of the conditions apply the CASE sequence will be exited without

any further commands being carried out and the method run continued with

the next command.

Simple example

Command Parameters Parameters

ДДДДДДДДДДДДД ДДДДДДДДДДДДДДДДДДДДДДДД ДДДДДДДДДДДДДДДДДДДДДДДД

MEAS_PH* pH measurement

Sensor comb. glass Signal drift 50 mV/min

Meas. input A1 Equilibr.time auto s

Temperature auto øC

(CASE CM<=7

DET_PH* Titration with NaOH

Reagent NaOH Sensor comb. glass

Concentr. 0.1 mol/L Meas. input A1

Dos. drive A1

CASE CM>7

DET_PH* Titration with HCl

Reagent HCl Sensor comb. glass

Concentr. 0.1 Mol/L Meas. input A1

Dos. drive A2

)CASE

END

In this case the sample is titrated with different reagents depending on the

pH of the sample.

It is possible to interlace CASE sequences in up to 3 levels.

Command Parameters Parameters

ДДДДДДДДДДДДД ДДДДДДДДДДДДДДДДДДДДДДДД ДДДДДДДДДДДДДДДДДДДДДДДД

(CASE condition 1. level

... ... commands

CASE condition 1. level

... ... commands

(CASE condition 2. level

... ... commands 2. level

)CASE

... ... commands

CASE condition 1. level

... ... commands

)CASE

The previously

measured pH value

is evaluated here