8. Command reference

8. Command reference

726 Titroprocessor, Instructions for use

153

153

Stop crit

.

Switch-off mode

Drift | time | none

The stop criterion defines the behavior of the titrator

when the endpoint is reached. The automatic termina-

tion of the titration can be carried out according to dif-

ferent criteria:

Drift

The titration is stopped when (after the EP

has been reached) the volume drift falls

below the set value.

Time

After the EP has been reached the set time

is allowed to elapse before the titration is

stopped.

none

The titration will not be stopped automati-

cally. The sample solution will be kept at

the endpoint.

If a renewed addition is necessary before the stop crite-

rion is recognized then a further volume increment will

be added and the stop criterion again checked.

Stop drift

Volume drift for titration termination

1…20…999 µL/min

After the last addition (…and after the EP has been

reached) the stop drift must be undercut before the ti-

tration can be ended. This volume drift can be inter-

preted as being the dosing rate which is theoretically

required to keep the measured value of the sample so-

lution constant at the endpoint.

t(delay)

Delay before titration termination

0…9999 s

When the EP has been reached the set time is allowed

to elapse before the titration is ended. If the delay time

is set as '0 s' then the titration will be terminated imme-

diately the EP has been reached.

8.4.3 Curve display with SET

The settings for the curve display correspond to those for the KFT mode,

see page 166.