10 sample changer commands, 1 sample changer commands – Metrohm 726 Titroprocessor User Manual

Page 189

background image

8. Command reference

8. Command reference

726 Titroprocessor, Instructions for use

183

183

8.10 Sample changer commands

ЪДДДДДДДДДДД¿

³Titration >³

³Changer >³

³CAL ЪДДДДДДД¿

³MEAS_* ³

CHANGER

³

³DOS ³SAMPLE ³

³STIR_I ³MOVE_S ³

³CASE ³MOVE_B ³

³COMM ³LIFT ³

³AUX ³PUMP ³

АДДДДДДДДДД³STIR_C ³

³RESET_C³

³(OMOVE ³

³)OMOVE ³

³(CMOVE ³

³)CMOVE ³

АДДДДДДДЩ

The versatile commands for controlling one (or two) 717 Sample

changer(s) allow the comprehensive automation of even complex working

procedures. With the sequences (OMOVE…)OMOVE and

(CMOVE…)CMOVE sample series can be processed in a very individual

manner. Conditional processes can be realized with

(CASE…EXIT…)CASE structures.
The standard methods supplied on the flash card (SCsimple.mth) for sam-

ple changer control can be easily altered and adapted to meet different

requirements.
Stored sample changer methods can usually be easily transferred to other

726 Titroprocessors. However, care must be taken that the rack definitions

(including working and rotating lift positions, positions and names of the

special beakers, etc.) of the corresponding sample changer are carried out

correctly.

8.10.1 Sample changer commands

CHANGER

Basic sample changer settings

The CHANGER command sets the active (of two possi-
ble) 717 Sample changer and defines its basic settings.

The command can be programmed at any point of the

run so that it is possible to operate two different sample

changers within a single method. The EBus address

selector disk on the appropriate sample changer must

be set correctly: for the first changer to 1, for the second

changer to 2, see sample changer installation on page

24.

Command Parameters Parameters

ДДДДДДДДДДДДД ДДДДДДДДДДДДДДДДДДДДДДДД ДДДДДДДДДДДДДДДДДДДДДДДД

CHANGER

Selection 1 Rack name

Lift rate 1 25 Lift rate 2 25

Shift rate 20 Beaker test mode single

Shift direction auto On beaker error MOVE_S

Selection

Address of the active sample changer

1 | 2

The changer address defines which changer is active

and for which sample changer the following basic set-

tings will apply.

Rack name

Name of the required rack

8 characters

This entry is optional. If a rack name has been entered

then the use of the described rack will be forced, i.e.