5.3 Determinations

726 Titroprocessor, Instructions for use

84

84

5.3.3 System variables

A determination produces the following system variables which are available

for calculations :

MNR$$

mean value of the results R$$ in statistics memory

MNX$$

mean value of the intermediate results X$$ in statistics

memory

MNC##

mean value of the common variables C##

ID1

sample identification 1 (only numerical entries)

ID2

sample identification 2 (only numerical entries)

ID3

sample identification 3 (only numerical entries)

SS

sample size (without units)

DD

determination duration in s

$RN

sample number

$AS

autostart counter

$BC

buffer counter

5.3.4 Determination files

Data produced during a determination are stored in a file in the working

memory while the determination is proceeding. The name of the

determination file and (optionally) file comments belonging to it are created

automatically according to the settings on the configuration page. Various

masks with combinations of determination time, sample number or sample

identifier, etc. are available, see page 34f.

If the statistics function is switched off (see below) the file of the current

determination will always be kept in the working memory until a new

determination has been started.

After the start of a determination the determination data (results, curves,

etc.) of the previous determination are no longer contained in the working

memory. The current determination file in the working memory cannot be

deleted and remains in existence – just like the current method – even when

the instrument is switched off. An exception to this is the initialization of the

statistics function, see below.

Statisticsfunction
If the statistics function is switched on then the statistics counter limit

determines the number of the determination files which can be present in

the working memory (Statistics 'on' 1 of '10' => max. 10 files
in the working memory). Up to 20 determination files can be contained in

the working memory.

If during a sample series the statistics counter becomes filled up with the

number of determinations carried out then at the start of the next (surplus)

determination the statistics will automatically be initialized again. This means

that the working memory will be completely emptied and can start to be

filled up again with the following determinations.

The statistics can be initialized manually and in this way the working

memory (together with the current determination) can be totally cleared. This

process is recommended before a sample series is carried out. Select
Statistics 'rst'

on the main page (this is not possible in the status

window).