beautypg.com

Metrohm 726 Titroprocessor User Manual

Page 47

background image

3. Configuration

726 Titroprocessor, Instructions for use

41

41

[Delete

all]

deletes all rack definitions.

[Check

code]

carries out a short test to check whether a rack code has
been allocated more than once. If this is the case then a

corresponding error message will be displayed.

3.2.2 Special beakers

Special beakers are the reserved beaker positions of a sample rack. Up to 8

special beakers can be defined per rack. In a method run they can be

placed in front of a tower for particular operations without interrupting

hindering the sample series. Special beakers can be used for rinsing the

electrode in a sequence of samples or to calibrate a sensor in an (OMOVE
start sequence (buffer solutions).

Special beakers are moved in front of tower 1 with the following command.

MOVE_B Special beaker spec1 to tower 1

Reserved special beaker positions, which can be defined separately for

each rack, are recognized as such in a method run and skipped when the

individual sample beakers are being processed.

If a special beaker is required in a method run but no special beaker is

present in the reserved position at the sample changer then an error

message is always displayed.

The softkey [Special beakers] opens the dialog window in which the
special beaker positions in the selected sample rack can be edited.

ЪДДД

Rack 1

ДДДДДДДДДДДДД¿

³ ³

³ # Spec.beaker Position ³

³ ДДДДДДДДДДДДДДДДДДДДДД ³

³ 1 Rinse 12 ³

³ 2 Cond. 11 ³

³ 3 spec.3 0 ³

³ 4 spec.4 0 ³

³ 5 spec.5 0 ³

³ 6 spec.6 0 ³

³ 7 spec.7 0 ³

³ 8 spec.8 0 ³

АДДДДДДДДДДДДДДДДД QUIT ÄÙ

Spec.beaker

Freely selectable name of the special beaker.

Position

Rack position of the special beaker (0=not defined)
It is recommended that special beakers are not placed

in the first rack positions but in higher rack positions.