2 configuration of sample changers, 1 sample racks, 2 special beakers – Metrohm 726 Titroprocessor User Manual
Page 43: 3 basic configuration of 717 sample changer
3. Configuration
726 Titroprocessor, Instructions for use
37
37
It is recommended that all the files in the internal
memory as well as the configuration settings (see
softkey [File]) are stored on data cards or on a
PC before the diagnosis function is carried out.
[>> >>]
switches to the second softkey bar.
3.2 Configuration of sample changers
In order to guarantee reliable sample changer operation it is important that
the settings of the connected sample changer 717 and the sample racks
used are configured correctly.
By pressing the softkey [Config.] and then the softkey [Sample
changer]
the configuration of the sample changer can be accessed from
the main page.
ÚÄÄÄConfiguration ДДДДДДДДДДДДДДДДДДДДДДДДД(Program version 5.726.0012)ДДДДДДДД¿
³ ЪДДД
Configuration: Sample changer
ДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДД¿ ³
³ Dial³ ³ ³
³ ³ Rack definitions ³ ³
³ LCD ³ # Name Beakers Special positions Type Code ³ ³
³ LCD ³ Work Shift Rinse Spec. ³ ³
³ ³ ДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДД ³ ³
³ Fehl³ 1 12*250mL 12 0 mm 0 mm 0 mm 0 mm M12-0 000001 ³ ³
³ Edit³ 2 16*150mL 16 0 mm 0 mm 0 mm 0 mm M16-0 000010 ³ ³
³ Edit³ 3 24*75mL 24 0 mm 0 mm 0 mm 0 mm M24-0 001000 ³ ³
³ ³ 6 12*150mL 12 0 mm 0 mm 0 mm 0 mm M12-0 100000 ³ ³
³ ³ 5 14*8oz 14 0 mm 0 mm 0 mm 0 mm M14-0 000101 ³ ³
³ ³ 4 14*200mL 14 0 mm 0 mm 0 mm 0 mm M14-0 000011 ³ ³
³ ³ 7 19*100mL 19 0 mm 0 mm 0 mm 0 mm M19-0 001001 ³ ³
³ ³ 8 0 mm 0 mm 0 mm 0 mm ³ ³
³ ³ 9 0 mm 0 mm 0 mm 0 mm ³ ³
³ ³ 10 0 mm 0 mm 0 mm 0 mm ³ ³
³ ³ 11 0 mm 0 mm 0 mm 0 mm ³ ³
³ ³ 12 0 mm 0 mm 0 mm 0 mm ³ ³
³ ³ 13 0 mm 0 mm 0 mm 0 mm ³ ³
³ ³ 14 0 mm 0 mm 0 mm 0 mm ³ ³
³ ³ 15 0 mm 0 mm 0 mm 0 mm ³ ³
³ ³ 16 0 mm 0 mm 0 mm 0 mm ³ ³
АДДДДДАДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДД QUIT ÄÙIT ДЩ
ЪДДДДДДДВДДДДДДДВДДДДДДД¿ ЪДДДДДДДВДДДДДДДВДДДДДДД¿ ЪДДДДДДДВДДДДДДДВДДДДДДД¿
³Delete ³ Rack ³ ³ ³Special³Changer³Changer³ ³Delete ³Delete ³ Check ³
³type ³ File ³ ³ ³beakers³ 1 ³ 2 ³ ³line ³ all ³ code ³
АДДДДДДДБДДДДДДДБДДДДДДДЩ АДДДДДДДБДДДДДДДБДДДДДДДЩ АДДДДДДДБДДДДДДДБДДДДДДДЩ