beautypg.com

Metrohm 726 Titroprocessor User Manual

Page 178

background image

8.8 MEAS

726 Titroprocessor, Instructions for use

172

172

The measuring parameters

MEAS_PH*

potent. or volta-/amperometric

measurement

Sensor Signal drift 50 mV/min

Meas. input A1 Equilibr.time auto s
Temperature 25.0

°C

applies to

MEAS_PH*, MEAS_U*, MEAS_UPOL*, MEAS_IPOL*

MEAS_T*

see below.

The measuring parameters apply to single, drift-controlled measurements.

Comment

Comments on measuring mode

(50 characters)

The commentary on the mode is used for identifying the

specific measurement within a determination. This is

chiefly used for more complex methods with several ti-

tration or measuring modes.

Signal drift

Limit for measurement drift

for MEAS_PH/U/IPOL:

0.5…50…999 mV/min | off

for MEAS_UPOL:

0.1…50…100 µA/min | off

The measured value is only accepted when the variation

in the sensor measurement falls below the given value

for the measurement drift. If this is delayed then the

measurement will nevertheless be accepted when the

preset delay period (see below) has elapsed.

This measured value drift control can be switched off

('off'). If drift control is switched off the preset delay
period (see below) will be strictly observed. This can be

favorable for multiple measurements with short meas-

uring interval cycles.

Meas. input

Sensor connection sockets

(A1,A2, Apol, Adiff, B1, B2, Bpol, Bdiff)

The measuring inputs A1, A2 and Apol are located on

the rear panel of the instrument in the measuring group

Sensors A. Adiff stands for the measuring amplifier

switch on the measurement inputs A1, A2 and Ref.

If a second measuring group (Sensors B) is present

then the addresses B1, B2 , Bpol and Bdiff can be

similarly used. Depending on the measuring quantity

only sensible sensor addresses can be selected,

e.g. for MEAS_IPOL* only Apol or Bpol.