12.1 Error messages
Error messages are accompanied by the name of the instrument module in
which the fault has occurred. In addition, each error message is provided
with a number so that it is easier to find the corresponding error message in
the following list.
A/D CONV
A/D CONV 1:
XX not corr. sensor for DEVICES and METHOD
In the method being carried out a sensor has been laid down which does
not match the current entry on the dialog page 'Devices & Manual Control'.
Check whether the correct sensor is being used. If necessary bring the entry
up to date (
names is not required this can be switched off by completely deleting the
entry.
A/D CONV 2:
X: check temperature sensor
Either no temperature sensor has been connected to the sensor interface X,
or it may have a fault.
A/D CONV 3:
interface X busy
•
The sensor to be accessed is currently being used for an actual deter-
mination. Wait until this has been completed.
•
A method in progress is trying to access a sensor which is currently be-
ing used for a manual measurement. This causes the determination to
be interrupted ('HOLD'). It can be resumed by pressing the CONT> key as soon as the affected sensor is available again. A/D CONV 4: interface X: unexpected error A fault has occurred which cannot be specified more closely. Please con- tact the service department of your local Metrohm agency and tell them the error message number. A/D CONV 5: interface X not present An attempt has been made to access a sensor interface which has not been installed. Check the method.