12 Troubleshooting

12.1 Error messages

Error messages are accompanied by the name of the instrument module in

which the fault has occurred. In addition, each error message is provided

with a number so that it is easier to find the corresponding error message in

the following list.

A/D CONV

A/D CONV 1:

XX not corr. sensor for DEVICES and METHOD

In the method being carried out a sensor has been laid down which does

not match the current entry on the dialog page 'Devices & Manual Control'.

Check whether the correct sensor is being used. If necessary bring the entry

up to date ( key). If the comparison of the sensor

names is not required this can be switched off by completely deleting the

entry.

A/D CONV 2:

X: check temperature sensor

Either no temperature sensor has been connected to the sensor interface X,

or it may have a fault.

A/D CONV 3:

interface X busy

The sensor to be accessed is currently being used for an actual deter-

mination. Wait until this has been completed.

A method in progress is trying to access a sensor which is currently be-

ing used for a manual measurement. This causes the determination to

be interrupted ('HOLD'). It can be resumed by pressing the

CONT> key as soon as the affected sensor is available again.

A/D CONV 4:

interface X: unexpected error

A fault has occurred which cannot be specified more closely. Please con-

tact the service department of your local Metrohm agency and tell them the

error message number.

A/D CONV 5:

interface X not present

An attempt has been made to access a sensor interface which has not been

installed. Check the method.