5. Methods and determinations

726 Titroprocessor, Instructions for use

79

79

5.3 Determinations

A method can be used to carry out a determination on a sample which can

include up to five titration or measurement modes. Each of these modes

can contain separate calculation formulas and report definitions. Each

mode produces its own data records (measuring point lists). These form the

basis of all reports and graphs. The complete report of a determination

comprises determination reports together with reports of the individual

titrations or measurements.

The data produced during a determination are stored in a file which can be

stored automatically in the internal memory or on a data card. It is also

possible to transfer the determination data directly via an RS232 interface as

a raw data report to a suitable PC database such as Metrodata VESUV 3 or

to store the whole file on a PC. Determination files can be comfortably

copied, renamed or deleted with the file manager.

Existing determination data can be subsequently re-evaluated and

recalculated without any problems and without having to load the original

method.

5.3.1 Carrying out a determination

A determination is started with the key. The defined start report is

printed out first.

The method is then processed line by line. As soon as an END command

is reached the determination will be terminated properly. The determination

data are stored in a file and the final report is printed. If the autostart

function has been switched on the next determination will be started;

otherwise the Titroprocessor 726 will return to the normal state.

A determination run can be interrupted with the key. The

determination can be continued by pressing the key

again. In the 'Hold' condition the Titroprocessor can be operated normally.

The ('Live') parameters of the method can be edited. The various peripheral

devices can be operated manually provided that they not blocked by the

determination. This is a particular advantage when a sample changer is

being used. The determination can also be completely terminated in the

'Hold' condition with the key.

A running determination can be terminated at any point with the

key. The necessary termination operations are carried out just as

in a regular determination. The type of determination termination is

mentioned in the results report (e.g. termination condition 'manual' or

'regular').
If the autostart function is switched on pressing the key will

terminate the complete series of samples.

If in a determination run containing several commands the execution of the

currently active command is to be interrupted and a jump made to the

following command in the method then this can be carried out with the

key by pressing the key combination

с

>.

Warning: during the start phase of a titration only the addition of the initial

volume is terminated with the key. The titration will nevertheless be

carried out normally.

START

HOLD
CONT

STOP

SKIP

HOLD
CONT