Manual operation

726 Titroprocessor, Instructions for use

55

55

well as the outlet port (for Dosinos) through which the reagent will be

expelled in this process, see page 46ff.

4.2.3 Manual dosing

In order to operate a Dosimat or a Dosino manually place the field cursor in

the column of the required dosing device and press the [Dosing drive]
softkey.

ЪДДД

Dosing

ДДДДДДДДДДДДДДДД¿

³ ³

³ Dos. drive : A1 ³

³ ³

³ Volume 50.00 mL ³

³ Dos. rate max mL/min ³

³ Fill. rate max mL/min ³

³ Pist. pos. 0.000 mL ³

³ ³

³ * 0.000 mL * ³

³ ³

³ ³

АДДДДДДДДДДДДДДДДДДДД QUIT ДЩ

ДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДД

ДДВДДДДДДД¿ ЪДДДДДДДВДДДДДДДВДДДДДДД¿

³ Fill ³ ³ HOLD ³Start ³ STOP ³

³ ³ ³ CONT ³dosing ³ ³

ДДБДДДДДДДЩ АДДДДДДДБДДДДДДДБДДДДДДДЩ

Enter the reagent volume to be dosed in. Alter the dosing and filling speeds

if necessary.

Softkeys

[Fill]

fills the buret.

[HOLD

CONT]

interrupts or restarts the dosing process.

[Start

dosing]

doses in the preselected volume.

[STOP]

breaks off the dosing process.