9. Results and statistics

726 Titroprocessor, Instructions for use

213

213

In the results window all necessary alterations for re-evaluations can be

carried out.

If several titration and measuring modes have been carried out in a single

determination then their results are listed underneath each other.

If the parameters have to be altered then the data record involved must first

be selected with the bar cursor.

Example:

ЪДДД Results: 97121011.011 Original data ДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДД¿

³ # 1 MEAS_PH* pH measurement ³

³ IM pH 4.356 pH value 4.36 pH ³

³ CM pH 4.357 ³

³ ³

³ # 2 DET_PH* Free Acid ³

³ EP1 pH 7.110 4.727 mL acid content 55.2 mol/L ³

³ ³

³ Error list: ³

³ ДДДДДДДДДДД ³

³ ³

³ ³

³ ³

³ ³

³ ³

³ ³

³ ³

³ ³

³ ³

³ ³

³ ³

³ ³

³ ³

АДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДД QUIT ДЩ

ЪДДДДДДДВДДДДДДДВДДДДДДД¿ ЪДДДДДДДВДДДДДДДВДДДДДДД¿ ЪДДДДДДДВДДДДДДДВДДДДДДД¿

³Sample ³Select ³Determ.³ ³ * ³ * ³ * ³ ³Local ³Calib. ³ >> ³

³data ³ ³report ³ ³ Param ³Formula³ Report³ ³ComVars³data ³ >> ³

АДДДДДДДБДДДДДДДБДДДДДДДЩ АДДДДДДДБДДДДДДДБДДДДДДДЩ АДДДДДДДБДДДДДДДБДДДДДДДЩ

Softkeys

[Sample

data]

opens an entry window for altering the sample data.

ЪДДД

Sample data

ДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДД¿

³ ³

³ User Run number 001 ³

³ ³

³ Method Ident1 Ident2 Ident3 Size Unit ³

³ ³

³ Remark ³

³ ³

³ ³

АДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДД QUIT ÄÙ

[Select]

opens a selection list with the determinations contained in
the working memory. When one of these is selected its

results are displayed.

[Determ.

report]

opens the dialog window for altering the determination
report, see page 74ff.