8. Command reference

8. Command reference

726 Titroprocessor, Instructions for use

199

199

8.13 Communication functions

³COMM >³

³AUX ЪДДДДДДДД¿

АДДДДДДДД³

SCAN_RM

³

³CTRL_RM ³

³CTRL_RS1³

³CTRL_RS2³

³SCAN_RS1³

³SCAN_RS2³

³PRINT ³

³KEYCTRL ³

АДДДДДДДДЩ

Complex applications require the optimal cooperation between various

instruments and components. For this reason the 726 Titroprocessor is

provided with special commands. These commands allow synchronization

and data communication between instruments which have a parallel remote

connection or a serial RS 232 interface.

The two commands CTRL and SCAN are important for the cooperation of

the 726 Titroprocessor with other instruments. They can also be used in

combination with instruments from other manufacturers provided that the

interface specifications of these are known.

8.13.1 Application mode of operation of CTRL and SCAN

The CTRL command sends the preset data without delay via the selected

RS 232 interface to a connected instrument or sets the predefined condi-

tions of the output lines (Output 0…7 or …13) of the selected remote con-

nection.

The SCAN command stops the method run and checks (scans) the in-

coming data at one of the two RS 232 interfaces or the conditions/signals of

the input lines (Input 0…7) of a remote connection. The method run will be

continued as soon as the defined data have been received or the incoming

signals correspond to those set down.

8.13.2 Remote connections

Most Metrohm instruments have a remote connection (D-sub, 25-pins),

which is provided for external control of the instrument. By setting particular

input lines various functions (start, stop etc.) can be triggered or different

measuring modes activated. At the same time the instrument signalizes its

instrument status or status change or certain events by activating its various

output lines.

The remote functions most often required for Metrohm instruments:

Function

Line

Status

Start

Input 0

active

Stop

Input 1

active

The status signals most often required for Metrohm instruments:

Signal

Line

Status

Ready

Output 0

active

End of Determination

Output 3

active (impulse)

Error

Output 5

active