8.5 SEC

726 Titroprocessor, Instructions for use

156

156

Adjustment

Predefined parameter sets

slow | medium | fast

A selection can be made from three different sets of pa-

rameters for the simplified setting of the titration parame-

ters. Each of these sets includes proven controller settings

which are suitable for most requirements.

slow

for extremely accurate titrations or for chemically

inhibited reactions requiring a lot of time

medium standard parameters for unproblematic

titrations, optimized for the commonest

applications

fast

speed-optimized control parameters

for non-critical titrations

If the titration parameters have to be adapted for specific ti-

tration problems the different control parameters can be

altered individually.

The [Custom adjust] softkey opens a further dialog
window in which user-defined settings can be carried out

(see page 151).

Fill. rate

Buret filling speed

0.01 mL/min…150 mL/min | max.

At the end of a titration the buret is automatically filled. The

maximum value for the filling rate depends on the cylinder

size of the buret used.

Buret cylinder max. speed

1 mL

3 mL/min

2 mL

6 mL/min

5 mL

15 mL/min

10 mL

30 mL/min

20 mL

60 mL/min

50 mL

150 mL/min

A value up to 150 mL/min can also be entered for smaller

buret cylinders; however, when the method is carried out

the dosing speed will automatically be reduced to the larg-

est possible value. Whenever possible 'max' should be se-

lected, except for viscous reagent solutions for which too

rapid filling can cause a partial vacuum in the cylinder

(bubble formation!).

Temperature

Temperature setting or measuring mode

-170…25.0…500°C | auto | initial

The temperature can be automatically determined during a

titration with the aid of a Pt100/Pt1000 sensor or a com-

bined electrode with built-in temperature sensor and can be

used for automatic pH correction (if pH is being measured).

It is possible to record the temperature throughout a titra-