Page 62: 3 calibration

background image

4.3 Sensors

726 Titroprocessor, Instructions for use

56

56

4.3 Sensors

Depending on the Titroprocessor model, two measuring groups could be

built in. The connected sensors can be operated from the 'Manual control

page'. A manual measurement is also possible while a determination is

being carried out. The sensors in use are blocked for manual operation by

the Titroprocessor. Manual measurements can only be carried out with one

sensor at a time.

A name can be entered for each sensor or measuring input.

Softkeys

[Calibr.]

opens the dialog window for the calibration of pH
electrodes.

[Meas]

opens the dialog window for manual measurements.

The required sensor (A1, A2, Adiff, AT or Apol) is first selected with the

cursor keys before a softkey is pressed. The tab key is used to access the

sensors of measuring group B.

4.3.1 Automatic sensor check

In order to ensure that sensor name in a method coincides with the current

settings of the manual control page (in which the actual occupancy of the

measuring inputs should be entered) it is possible to force a comparative

check to be carried out.

If a sensor name has been entered in the method and in the 'Manual control

page' then an automatic sensor check is carried out. The two entries are

compared with each other. If they do not agree an error message is

produced at the start of the particular mode.

The automatic sensor check is suppressed if no entries have been made in

the method or if the corresponding entries have been completely deleted in

the 'Manual control page'.

³ Meas. input A1

³ Sensor 6.0232.100

ß Manual control page

â Method

DET_PH* Determination of chloride

Reagent AgNO3 Sensor 6.0232.100

Concentr. 0.01 mol/L Meas. input A1

Dos. drive A1

4.3.2 Manual measurement

Measurements can be carried out by selecting the measuring input to which

the corresponding sensor is connected. The [Meas] softkey opens a
dialog window in which the measurement parameters, e.g. measuring mode

or measuring temperature can be modified according to the type of sensor.