3. Configuration

726 Titroprocessor, Instructions for use

33

33

3 Configuration

In order to guarantee problem-free work with the 726 Titroprocessor it is

essential that the basic settings of all instrument components are

configured according to your requirements.

3.1 Basic settings

By pressing the softkey [Config.] the basic configuration can be
accessed from the main page.

ÚÄÄÄConfiguration ДДДДДДДДДДДДДДДДДДДДДДДДД(Program version 5.726.0012)ДДДДДДДД¿

³ ³

³ Dialog in english Date 1997-12-04 ³

³ Time 16:05:52 ³

³ LCD contrast 100 ³

³ Screen save 30 min Standard data file name DATA ³

³ File name type .### ³

³ Error beep on File comment type Ident1,2+Meth+Remark ³

³ Edit beep on ³

³ Edit mode insert Power on PREP warning on ³

³ ³

³ Remote terminal mode off ³

³ File transfer protocol Metrohm(Bkup) ³

³ ³

³ ³

³ ³

³ ³

³ ³

³ ³

³ ³

³ ³

³ ³

³ ³

АДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДДД QUIT ДЩ

ЪДДДДДДДВДДДДДДДВДДДДДДД¿ ЪДДДДДДДВДДДДДДДВДДДДДДД¿ ЪДДДДДДДВДДДДДДДВДДДДДДД¿

³Device ³File ³Access ³ ³Sample ³Inter- ³Config.³ ³Tubings³ ³ >> ³

³header ³ ³control³ ³changer³faces ³idents.³ ³ ³ ³ >> ³

АДДДДДДДБДДДДДДДБДДДДДДДЩ АДДДДДДДБДДДДДДДБДДДДДДДЩ АДДДДДДДБДДДДДДДБДДДДДДДЩ

The basic settings for the Titroprocessor 726 can be configured on this

dialog page. The softkeys allow access to the configuration of further

components.

Dialog in

Dialog language. English, German, French and
Spanish are available. After changing the dialog

language the instrument should be switched off and

then back on again.

LCD contrast

Contrastcan be selected in 250 steps.
The contrast can also be set with the key

combinations

й

> and

к

>.

Screen save

Delay period for switching the screen off. If no key
on the Titroprocessor is pressed within the set period

then the background illumination of the LCD screen

will be automatically switched off (for its own care).

Pressing a key switches it on again.

Error beep

Acoustic error message signal on/off

Edit beep

Acoustic editing error message signal on/off

Edit mode

Switches between overwrite/insert mode during data