beautypg.com

Metrohm 726 Titroprocessor User Manual

Page 125

background image

7. Titration and measuring modes

726 Titroprocessor, Instructions for use

119

119

Remark

Formula name

11 characters

Each formula can be given any name. The name only has

a descriptive character.

Formula

Calculation formula

35 characters

A formula consists of a variable (final result, intermediate

result or common variable), which is allocated to the re-

sult of the calculation and the calculation formula itself,

which can consist of mathematical operators, numerical

values, common variables and system variables. Results

of previous calculations also belong to the system vari-

ables.

Calculation results can be allocated the following vari-

ables:

R00…RZZ

Final result, is given on the result page
and in reports; can be used for calcula-

tions, statistics and live displays.

X00…XZZ

Intermediate result, is not shown; can
be used for calculations, statistics and

live displays.

C01…C19

Common variables, are given on the
result page and in reports; can be used

for calculations, statistics and live dis-

plays. They are globally available for

subsequent determinations.

Unit

Unit of the result

% | ppm | g/L | mg/mL | mg/pc |

mol/L | g | mg | mL | s | mL/min |

max. 6 characters

Any other unit can be entered apart from the selection of

units given above,.

Dec

Decimal place in the result

0…5

The result of the calculation will be rounded off to the

number of decimal places given above and this rounded-

off value enters the statistics. If the result is used internally

for other calculations then the original and completely ac-

curate value will be used.

Type

Type of result

normal | stat. | live

The type of result determines how it will be shown:

normal

The result is shown in the result
window and in the report.